Section 228.74. Absorbed dose rate  


Latest version.
  • New equipment shall have a system that provides information from which the absorbed dose rate at a reference point in the treatment volume can be calculated. The radiation detectors specified in § 228.66 (relating to beam monitors) may form part of this system. The dose monitor unit rate shall be displayed at the control panel.