Section 225.92. Radiation survey meter calibration requirements
-
(a) In addition to the requirements of § 225.91 (relating to survey meter requirements), instruments required by this chapter shall have a range so that 0.516 µC/kg (2 mR) per hour through 258 µc/kg (1 R) per hour can be measured.
(b) Each radiation instrument shall be calibrated:
(1) At energies appropriate for use.
(2) At intervals not to exceed 6 months.
(3) After each instrument servicing, other than battery replacement.
(4) To within an accuracy of +/- 20%.
(5) At two points located approximately one-third and two-thirds of full scale on each scale of linear scale instruments; at mid-range of each decade and at two points of at least 1 decade for logarithmic scale instruments; and for digital instruments, at three points between 0.516 µC/kg (2 mR) and 258 µC/kg (1000 mR) per hour.
(6) By a person authorized by the Department, the NRC or an agreement state.
(c) Calibration records shall be maintained for inspection by the Department for 3 years after the date of calibration.