Pennsylvania Code (Last Updated: April 5, 2016) |
Title 25. ENVIRONMENTAL PROTECTION |
PART I. Department of Environmental Protection |
Subpart D. Environmental Health and Safety |
Article V. Radiological Health |
Chapter 221. X-rays in the Healing Arts |
Table II.
-
Design operating range (Kilovolts peak) Measured potential (Kilovolts peak) Minimum half-value layer
(millimeters of aluminum)Specified dental systems* All other X-ray systems Below 51 30 1.5 0.3 40 1.5 0.4 50 1.5 0.5 51 to 70 51 1.5 1.2 60 1.5 1.3 70 1.5 1.5 Above 70 71 2.1 2.1 80 2.3 2.3 90 2.5 2.5 100 2.7 2.7 110 3.0 3.0 120 3.2 3.2 130 3.5 3.5 140 3.8 3.8 150 4.1 4.1 Note: Half-value layers for kilovoltages not listed in Table II may be determined by interpolation or extrapolation.
* Dental systems manufactured after December 1, 1980, designed for use with intraoral image receptors.
(b) Beryllium window tubes shall have a minimum of 0.5 millimeter aluminum equivalent filtration permanently installed in the useful beam.
(c) For capacitor energy storage equipment, compliance with this section shall be determined with the maximum quantity of charge per exposure.
(d) The required minimal aluminum equivalent filtration shall include the filtration contributed by materials which are always present between the source and the patient.
(e) For X-ray systems having variable filtration in the useful beam, a means shall be provided to prohibit exposure unless the filtration requirements of subsection (a) are met for the kVp selected.
Authority The provisions of this § 221.25 issued under section 302 of the Radiation Protection Act (35 P. S. § 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20); amended under sections 301 and 302 of the Radiation Protection Act (35 P. S. § § 7110.301 and 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20).
Source The provisions of this § 221.25 adopted December 18, 1987, effective December 19, 1987, 17 Pa.B. 5235; amended July 16, 2004, effective July 17, 2004, 34 Pa.B. 3238. Immediately preceeding text appears at serial pages (249290) and (285671).
Cross References This section cited in 25 Pa. Code § 221.35a (relating to fluoroscopic X-ray systems).