Section 221.41a. Fluoroscopic timer  


Latest version.
  • A cumulative timing device activated by the fluoroscope switch shall be provided. It shall indicate the passage of a predetermined period of irradiation either by an audible signal or by temporary or permanent interruption of the irradiation when the increment of exposure time exceeds a predetermined limit not exceeding 5 minutes.

The provisions of this § 221.41a adopted October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894.