Section 221.39a. Barrier transmitted radiation rate limits  


Latest version.
  • The protective barrier may not transmit more than 2 milliroentgens (.516 µmC/kg) per hour at 10 centimeters from an accessible surface of the fluoroscopic imaging assembly for each roentgen per minute of entrance exposure rate.

The provisions of this § 221.39a adopted October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894.