Section 221.29. Kilovoltage (kV) accuracy
-
(a) For variable kV units, the kV output may not vary from the set-indicated value by more than 10% over the range of technique factors normally used. Discrepancies of more than 10% between set-indicated and measured kV values shall be investigated by a qualified expert or service agent and appropriate action taken.
(b) For fixed kV units, the kV output may not vary from the set-indicated value by more than 20% over the range of technique factors normally used. Discrepancies of more than 20% between set-indicated and measured kV values shall be investigated by a qualified expert or service agent and appropriate action taken.
The provisions of this § 221.29 adopted October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894; amended November 16, 2001, effective November 17, 2001, 31 Pa.B. 6282; amended July 16, 2004, effective July 17, 2004, 34 Pa.B. 3823. Immediately preceding text appears at serial pages (285672) to (285673).