Section 221.23. Leakage radiation from the diagnostic source assembly  


Latest version.
  • The leakage radiation from the diagnostic source assembly measured at a distance of 1 meter in any direction from the source may not exceed 100 milliroentgens (25.8 µC/kg) in 1 hour when the X-ray tube is operated at its leakage technique factors. Compliance shall be determined by measurements averaged over an area of 100 square centimeters with no linear dimension greater than 20 centimeters.

The provisions of this § 221.23 adopted February 1, 1972, effective February 2, 1972, 2 Pa.B. 212; amended December 18, 1987, effective December 19, 1987, 17 Pa.B. 5235. Immediately preceding text appears at serial page (4882).